EchelonGraph verdictPlan mitigationNo fix on record since 2020: if the vendor still supports the product, apply the fixed version it names within your standard update timelines; if it no longer does, isolate the affected system on your normal schedule, or retire the product (the SSVC decision on this page).
- •High severity, but no confirmed exploitation yet
CISA-KEV: Not listedEPSS PROB: 3.8%CVSS: 8.8Exploit: None knownExposed services: Not assessed
No fix has been recorded since the CVE record was published in 2020 — if the vendor still supports the product, apply the fixed version it names; if not, isolate the affected system (restrict network access to it, for example with a firewall, or keep it off the internet) or retire the product.
Medtronic MyCareLink Smart 25000 is
vulnerable to a race condition in the MCL Smart Patient Reader software update system, which allows unsigned firmware to be uploaded and executed on the Patient Reader. If exploited, an attacker could remotely execute code on the MCL Smart Patient Reader device, leading to control of the device.
What the EG score is made of (scorer v4.0: the three questions the 0–10 score mixes, published apart · methodology)
The EG score (8.8): severity, escalated by confirmed exploitation. This CVE's 8.8 is its impact severity (EG Severity), not escalated.
- EG Severity
- 8.8 HIGHCVSS-B, impact only: no KEV listing, EPSS or EG-KEV tier raises it.NVD CVSS v3 · CVSS v3.0 or v3.1
- EG Likelihood
- 3.8% within 30 daysSource: FIRST EPSS (model v2026.06.15, score date 2026-10-28). FIRST's exploitation probability, not an EchelonGraph model.
- Confirmed exploitation
- None on record (CISA KEV, EG-KEV Tier 1)Reported apart from the likelihood: an observed fact, not a probability.
- EG Priority
- SSVC Track at medium mission impactTrack at low or medium mission impact; Track* at high (mission-essential systems).CISA BOD 26-04: 14 days if the affected system is publicly exposed; Fix on system upgrade if it is not.No released fix is confirmed on this page, but CISA's timeline does not wait for a fix's release: it starts when an agency identifies the flaw on an asset and updates the CDM Agency and Federal Dashboard, and the directive counts patching, decommissioning the system or another action as remediation. Until a fixed release is applied, mitigate as this page's verdict says. In the directive's words, “Fix on system upgrade” means that, “unless conditions change”, the vulnerability “should be remediated the next time the vulnerable asset receives a scheduled major upgrade or rebuild”: its timelines are “dynamic as facts change”, and a KEV listing shortens them.Each input and its source, and whom BOD 26-04 binds: the SSVC and BOD 26-04 blocks below.
SSVC (CISA's decision table, applied by EchelonGraph)Track at low or medium mission impact; Track* at high (mission-essential systems).
No fix has been recorded since the CVE record was published in 2020. If the vendor still supports the product, apply the fixed version it names within your standard update timelines; if it no longer does, isolate the affected system (restrict network access to it, for example with a firewall, or keep it off the internet) on your normal schedule, or retire the product.
Exploitation none (no CISA value, KEV listing or exploit code on record) · Automatable no (CISA's BOD 26-04 default) · Technical impact total (CISA's BOD 26-04 default). EchelonGraph holds no SSVC assessment of this CVE from CISA: none of these inputs is from Vulnrichment. EchelonGraph applied CISA's decision table to them; CISA publishes SSVC inputs and the table, not a decision for each CVE. Mission impact is CISA's Mission & Well-being decision point, and only you can judge it. CISA's table makes it high when Mission Prevalence is Essential — the vulnerable component "directly provides capabilities that constitute at least one MEF for at least one entity" (MEF: mission essential function) — or when Public Well-Being Impact is Irreversible: "multiple fatalities are likely", the cyber-physical system "is likely lost or destroyed", "extreme or serious externalities" are imposed on other parties, or social systems such as elections or the financial grid "are destabilized and potentially collapse". CISA's decision table
BOD 26-04 (CISA's remediation timeline, Table 1)14 days if the affected system is publicly exposed; Fix on system upgrade if it is not.
No released fix is confirmed on this page, but CISA's timeline does not wait for a fix's release: it starts when an agency identifies the flaw on an asset and updates the CDM Agency and Federal Dashboard, and the directive counts patching, decommissioning the system or another action as remediation. Until a fixed release is applied, mitigate as this page's verdict says. In the directive's words, “Fix on system upgrade” means that, “unless conditions change”, the vulnerability “should be remediated the next time the vulnerable asset receives a scheduled major upgrade or rebuild”: its timelines are “dynamic as facts change”, and a KEV listing shortens them.
The score's inputs: In CISA KEV no (CISA's KEV catalog) · Automatable no (CISA's BOD 26-04 default) · Technical impact total (CISA's BOD 26-04 default). Where an input reads “CISA's BOD 26-04 default”, no one has published it for this CVE, and CISA's implementation guidance says: “If “Automatable” and “Technical Impact” information items about a CVE ID are not available and the CVE ID is not in the KEV Catalog, CISA will treat its values as “no” and “total,” respectively, until CISA or a third party provides metadata about the CVE ID. CISA will always provide the Vulnrichment data for a CVE ID listed in the KEV Catalog.” Whether your system is publicly exposed is yours to answer. BOD 26-04 binds US Federal Civilian Executive Branch agencies; for anyone else it is CISA's published timeline, a reference rather than an obligation. The directive · how EchelonGraph applies it